Operating menu
Micropilot NMR84
158
Hauser
V1 address
Navigation
Setup → Advanced setup → Communication → V1 X1-4 → Configuration → V1
address
Prerequisite
Communication interface protocol variant (→ 157)
Description
Identifier of the previous device for V1 communication.
User entry
0 to 255
Factory setting
1
Additional information
Read access
Operator
Write access
Maintenance
Level mapping
Navigation
Setup → Advanced setup → Communication → V1 X1-4 → Configuration → Level
mapping
Prerequisite
Communication interface protocol (→ 153)
V1
Description
Determines the transmittable range of levels.
Selection
• +ve
• +ve & -ve
Factory setting
+ve
Additional information
Read access
Operator
Write access
Maintenance
In V1, the level is always represented by a number in the range from 0 to 999
999. This
number corresponds to a level as follows:
"Level mapping" = "+ve"
Number
Corresponding level
0
0.0 mm
999
999
99
999.9 mm
"Level mapping" = "+ve & -ve"
Number
Corresponding level
0
0.0 mm
500
000
50
000.0 mm
Содержание Micropilot NMR84
Страница 177: ...Micropilot NMR84 Operating menu Endress Hauser 177 Additional information Read access Operator Write access Maintenance ...
Страница 204: ...Operating menu Micropilot NMR84 204 Endress Hauser Additional information Read access Operator Write access Maintenance ...
Страница 219: ...Micropilot NMR84 Operating menu Endress Hauser 219 Additional information Read access Operator Write access Maintenance ...
Страница 238: ...Operating menu Micropilot NMR84 238 Endress Hauser Read access Operator Write access Maintenance ...
Страница 255: ......