Micropilot NMR84
Repair
Hauser
95
13
Repair
13.1 General information on repairs
13.1.1 Repair concept
The Hauser repair concept assumes that the devices have a modular design and
that repairs can be done by the Hauser service or specially trained customers.
Spare parts are contained in suitable kits. They contain the related replacement
instructions.
For more information on service and spare parts, contact the Service Department at
Hauser.
13.1.2 Repairs to Ex-approved devices
When carrying out repairs to Ex-approved devices, please note the following:
• Repairs to Ex-approved devices may only be carried out by trained personnel or by the
Hauser Service.
• Comply with the prevailing standards, national Ex-area regulations, safety instructions
(XA) and certificates.
• Only use original spare parts from Hauser.
• When ordering a spare part, please note the device designation on the nameplate. Only
replace parts with identical parts.
• Carry out repairs according to the instructions. On completion of repairs, carry out the
specified routine test on the device.
• Only Hauser Service may convert a certified device into a different certified
variant.
• Document all repair work and conversions.
13.1.3 Replacement of a device or electronic module
After a complete device or the electronic mainboard has been replaced, the parameters can
be downloaded into the instrument again via FieldCare.
Condition: The configuration of the old device has been saved to the computer via
FieldCare.
You can continue to measure without carrying out a new setup. Only a linearization and a
tank map (interference echo suppression) have to be recorded again.
The "Save/Restore" function
After a device configuration has been saved to a computer and restored to the device
using the
Save/Restore
function of FieldCare, the device must be restarted by the
following setting:
Setup
→
Advanced setup
→
Administration
→
Device reset
=
Restart device
.
This ensures correct operation of the device after the restore.
Содержание Micropilot NMR84
Страница 177: ...Micropilot NMR84 Operating menu Endress Hauser 177 Additional information Read access Operator Write access Maintenance ...
Страница 204: ...Operating menu Micropilot NMR84 204 Endress Hauser Additional information Read access Operator Write access Maintenance ...
Страница 219: ...Micropilot NMR84 Operating menu Endress Hauser 219 Additional information Read access Operator Write access Maintenance ...
Страница 238: ...Operating menu Micropilot NMR84 238 Endress Hauser Read access Operator Write access Maintenance ...
Страница 255: ......