Micropilot NMR84
Diagnostics and troubleshooting
Hauser
87
11.4 Overview of the diagnostic messages
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
Diagnostic of sensor
102
Sensor incompatible error
1. Restart device
2. Contact service
F
Alarm
150
Detector error
1. Restart device
2. Check electrical connections
of detector
3. Replace detector unit
F
Alarm
151
Sensor electronic failure
Replace sensor electronic
module
F
Alarm
Diagnostic of electronic
242
Software incompatible
1. Check software
2. Flash or change main
electronic module
F
Alarm
252
Modules incompatible
1. Check if correct electronic
modul is plugged
2. Replace electronic module
F
Alarm
261
Electronic modules
1. Restart device
2. Check electronic modules
3. Change I/O module or main
electronics
F
Alarm
262
Module connection
1. Check module connections
2. Change electronic modules
F
Alarm
270
Main electronics failure
Replace main electronics
F
Alarm
271
Main electronics failure
1. Restart device
2. Change main electronic
module
F
Alarm
272
Main electronics failure
Restart device
F
Alarm
272
Main electronics failure
1. Restart device
2. Contact service
F
Alarm
273
Main electronics failure
1. Emergency operation via
display
2. Change main electronics
F
Alarm
275
I/O module failure
1. Restart device
2. Change I/O module
F
Alarm
276
I/O module faulty
1. Restart device
2. Change I/O module
F
Alarm
282
Data storage
1. Restart device
2. Contact service
F
Alarm
283
Memory content
1. Transfer data or reset device
2. Contact service
F
Alarm
284
Detector SW update in
progress
Firmware update active, please
wait!
F
Alarm
311
Electronic failure
Maintenance required!
1. Do not perform reset
2. Contact service
M
Warning
333
System recovery required
HW change detected
System configuration recovery
required
Go to menu on device and
perform recovery
F
Alarm
Содержание Micropilot NMR84
Страница 177: ...Micropilot NMR84 Operating menu Endress Hauser 177 Additional information Read access Operator Write access Maintenance ...
Страница 204: ...Operating menu Micropilot NMR84 204 Endress Hauser Additional information Read access Operator Write access Maintenance ...
Страница 219: ...Micropilot NMR84 Operating menu Endress Hauser 219 Additional information Read access Operator Write access Maintenance ...
Страница 238: ...Operating menu Micropilot NMR84 238 Endress Hauser Read access Operator Write access Maintenance ...
Страница 255: ......
Страница 256: ...www addresses endress com 71545626 71545626 ...