Micropilot NMR84
Diagnostics and troubleshooting
Hauser
85
11.3 Diagnostic information in FieldCare
Any faults detected by the measuring device are displayed on the home page of the
operating tool once the connection has been established.
Xxxxxx/…/…/
Xxxxxx
P
P
P
+
+
+
+
Diagnostics 1:
Remedy information:
Failure (F)
Function check (C)
Out of spezification (S)
Maintenance required (M)
C485 Simulation measured vari...
Deactivate Simulation (Service...
Diagnostics 1:
C485 Simu...
Remedy information:
Deactivate...
Access status tooling:
Mainenance
Operation
Setup
Diagnostics
Expert
Function check (C)
Status signal:
Device tag:
Device name:
Mass flow:
Volume flow:
Xxxxxxx
12.34
kg/h
12.34
m /h
³
Xxxxxxx
1
2
3
A0045844
1
Status area with status signal
2
Diagnostic information
3
Remedial measures with Service ID
Furthermore, diagnostic events that have occurred can be viewed in the
Diagnostic
list
submenu.
11.3.1 Status signals
The status signals provide information on the state and reliability of the device by
categorizing the cause of the diagnostic information (diagnostic event).
Symbol
Meaning
A0017271
Failure
A device error has occurred. The measured value is no longer valid.
A0017278
Function check
The device is in service mode (e.g. during a simulation or a warning).
A0017277
Out of specification
The device is operated outside its technical specification limits (e.g. outside the process
temperature range)
A0017276
Maintenance required
Maintenance is required. The measured value is still valid.
The status signals are categorized in accordance with VDI/VDE 2650 and NAMUR
Recommendation NE 107.
Содержание Micropilot NMR84
Страница 177: ...Micropilot NMR84 Operating menu Endress Hauser 177 Additional information Read access Operator Write access Maintenance ...
Страница 204: ...Operating menu Micropilot NMR84 204 Endress Hauser Additional information Read access Operator Write access Maintenance ...
Страница 219: ...Micropilot NMR84 Operating menu Endress Hauser 219 Additional information Read access Operator Write access Maintenance ...
Страница 238: ...Operating menu Micropilot NMR84 238 Endress Hauser Read access Operator Write access Maintenance ...
Страница 255: ......
Страница 256: ...www addresses endress com 71545626 71545626 ...