Micropilot NMR84
Operating menu
Hauser
197
Additional information
Read access
Operator
Write access
Maintenance
The
Covered
option is only valid for fixed tank roofs. For a floating roof select
Open
top
.
Stilling well
Navigation
Setup → Advanced setup → Application → Tank calculation → CTSh → Stilling well
Description
Determines whether the device is mounted on a stilling well.
Selection
• No
• Yes
Factory setting
No
Additional information
Read access
Operator
Write access
Maintenance
Calibration temperature
Navigation
Setup → Advanced setup → Application → Tank calculation → CTSh → Calibration
temp
Description
Specify temperature at which the measurement has been calibrated.
User entry
–50 to 250 °C
Factory setting
25 °C
Additional information
Read access
Operator
Write access
Maintenance
Linear expansion coefficient
Navigation
Setup → Advanced setup → Application → Tank calculation → CTSh → Linear exp
coeff
Description
Defines the linear expansion coefficient of the tank shell material.
User entry
0 to 100 ppm
Factory setting
15 ppm
Содержание Micropilot NMR84
Страница 177: ...Micropilot NMR84 Operating menu Endress Hauser 177 Additional information Read access Operator Write access Maintenance ...
Страница 204: ...Operating menu Micropilot NMR84 204 Endress Hauser Additional information Read access Operator Write access Maintenance ...
Страница 219: ...Micropilot NMR84 Operating menu Endress Hauser 219 Additional information Read access Operator Write access Maintenance ...
Страница 238: ...Operating menu Micropilot NMR84 238 Endress Hauser Read access Operator Write access Maintenance ...
Страница 255: ......
Страница 256: ...www addresses endress com 71545626 71545626 ...