Micropilot NMR84
Operating menu
Hauser
217
Additional information
Read access
Operator
Write access
Maintenance
Alarm hysteresis
Navigation
Setup → Advanced setup → Application → Alarm → Alarm → Alarm hysteresis
Prerequisite
Alarm mode (→ 210)
≠
Off
Description
Defines the hysteresis for the limit values. The hystersis prevents constant changes of the
alarm state if the level is near one of the limit values.
User entry
Signed floating-point number
Factory setting
0.001
Additional information
Read access
Maintenance
Write access
Maintenance
Damping factor
Navigation
Setup → Advanced setup → Application → Alarm → Alarm → Damping factor
Description
Defines the damping constant (in seconds).
User entry
0 to 999.9 s
Factory setting
0 s
Additional information
Read access
Operator
Write access
Maintenance