Micropilot NMR84
Operating menu
Hauser
123
Meaning of the options
•
No
The map is not recorded.
•
Record map
The map is recorded. After the recording is completed, the new measured distance and
the new mapping range appear on the display. When operating via the local display,
these values must be confirmed by pressing .
•
Recalculate map
Used internally by the software. Initiates a new calculation of the map from the new
data points.
•
Overlay map
The new mapping curve is generated by overlaying the old and the current envelope
curves.
•
Factory map
The factory map stored in the ROM of the device is used.
•
Delete partial map
The mapping curve is deleted up to
•
Stop overlay
Stops the overlaying of the map.
Distance
Navigation
Setup → Distance
Description
Distance from lower edge of device flange to product surface.
Additional information
Read access
Operator
Write access
-
Liquid temp source
Navigation
Setup → Liq temp source
Description
Defines source from which the liquid temperature is obtained.
Selection
• Manual value
• HART device 1 ... 15 temperature
• AIO B1-3 value
• AIO C1-3 value
• AIP B4-8 value
• AIP C4-8 value
Factory setting
Manual value
Additional information
Read access
Operator
Write access
Maintenance
Содержание Micropilot NMR84
Страница 177: ...Micropilot NMR84 Operating menu Endress Hauser 177 Additional information Read access Operator Write access Maintenance ...
Страница 204: ...Operating menu Micropilot NMR84 204 Endress Hauser Additional information Read access Operator Write access Maintenance ...
Страница 219: ...Micropilot NMR84 Operating menu Endress Hauser 219 Additional information Read access Operator Write access Maintenance ...
Страница 238: ...Operating menu Micropilot NMR84 238 Endress Hauser Read access Operator Write access Maintenance ...
Страница 255: ......
Страница 256: ...www addresses endress com 71545626 71545626 ...