Deltabar S FMD77, FMD78, PMD75 with FOUNDATION Fieldbus
Commissioning
Hauser
69
7.6.3
Quick Setup menu for the Flow measuring mode (FieldCare)
P01-xxxxxxxx-19-xx-xx-en-166
Fig. 37:
Quick Setup menu for the "Flow" measuring mode
For onsite operation, see also
37, Section 6.2.3 "Function of the operating elements –
51, Section 6.4 "Local operation – onsite display
1
)
1
)
2
)
2
)
1
)
1
)
O
n
-
si
te
d
isplay
on
ly
2
)
Fi
e
l
d
Ca
re
on
ly
P
RO
FI
L
E
VI
E
W
M
AX
F
LO
W
M
AX
. P
R
E
SS
. F
LO
W
P
O
S
.
ZE
RO A
D
JU
S
T
D
A
MPIN
G
V
AL
U
E
M
E
A
S
U
R
E
D
V
AL
U
E
QUICK
S
E
T
UP
GRO
UP
S
E
L
E
C
T
I
O
N
M
E
A
S
U
R
IN
G
M
O
D
E
M
E
A
S
U
R
IN
G
M
O
D
E
LA
N
G
U
AG
E
Flow
P
re
ssu
re
L
e
v
e
l
QUICK
S
E
T
UP
O
P
E
RAT
IN
G
M
E
NU
Local operation
FieldCare
Measured value display
Onsite display: Switch from the measured value
display to GROUP SELECTION with F .
Measured value display
Select QUICK SETUP menu.
GROUP SELECTION
Select MEASURING MODE.
Measuring Mode
Select the Primary Value Type parameter.
MEASURING MODE
Select "Flow" option.
GROUP SELECTION
Select QUICK SETUP menu.
Primary Value Type
Select "Flow" option.
POS. ZERO ADJUST
Due to orientation of the device, there may be a shift
in the measured value. You correct the MEASURED
VALUE via the POS. ZERO ADJUST parameter with
the "Confirm" option, i.e. you assign the value 0.0 to
the pressure present.
POS. ZERO ADJUST
Due to orientation of the device, there may be a shift
in the measured value. You correct the MEASURED
VALUE via the POS. ZERO ADJUST parameter with
the "Confirm" option, i.e. you assign the value 0.0 to
the pressure present.
MAX. FLOW
Enter maximum flow of primary device.
(
See also layout sheet of primary device).
MAX. FLOW
Enter maximum flow of primary device.
(
See also layout sheet of primary device).
MAX. PRESS. FLOW
Enter maximum pressure of primary device.
(
See also layout sheet of primary device).
MAX. PRESS. FLOW
Enter maximum pressure of primary device.
(
See also layout sheet of primary device).
DAMPING VALUE
Enter damping time (time constant
). The damping
affects the speed at which all subsequent elements,
such as the onsite display, measured value and the
OUT Value of the Analog Input Block react to a
change in the pressure.
DAMPING VALUE
Enter damping time (time constant
). The damping
affects the speed at which all subsequent elements,
such as the onsite display, measured value and the
OUT Value of the Analog Input Block react to a
change in the pressure.