Deltabar S FMD77, FMD78, PMD75 with FOUNDATION Fieldbus
Operation
Hauser
55
5. Operation via an FF configuration program: using the DAT_HANDLING/ HistoROM
CONTROL parameter in the Service Transducer Block, select the "Device
HistoROM"
option for the data transfer direction.
Operation via FieldCare: using the HistoROM CONTROL parameter, select the "Device
HistoROM" option for the data transfer direction. (Menu path: OPERATING MENU
OPERATION).
Using the DOWNLOAD SELECT. parameter (OPERATION menu), select which
parameters should be overwritten.
The following parameters are overwritten depending on the option selected:
–
Configuration copy
:
All the parameters apart from the TRANSMITTER SERIAL NO., DEVICE
DESIGNATION and the parameters of the POSITION ADJUSTMENT and PROCESS
CONNECTION group.
–
Device replacement:
All the parameters apart from the TRANSMITTER SERIAL NO., DEVICE
DESIGNATION and the parameters of the POSITION ADJUSTMENT and PROCESS
CONNECTION group.
–
Electronics replace:
All parameters apart from the parameters of the POSITION ADJUSTMENT group
.
Factory setting: Configuration copy
6. Using the HistoROM CONTROL parameter, select the "Device
HistoROM" option for
the data transfer direction.
7. Wait approx. 40 seconds. Configuration data are uploaded from the device to the
HistoROM
®
/M-DAT. The device is not restarted.
8. Disconnect the device from the supply voltage again.
9. Detach the memory module.
10. Reestablish supply voltage to the device.
Copying configuration data from a HistoROM
®
/M-DAT to a device
:
Operation must be unlocked.
1. Disconnect the device from the supply voltage.
2. Attach the HistoROM
®
/M-DAT to the electronic insert. Configuration data from
another device are stored in the HistoROM
®
/M-DAT.
3. Reestablish supply voltage to the device.
4. Operation via an FF configuration program: using the DAT_HANDLING/ HistoROM
CONTROL parameter in the Service Transducer Block, select the "HistoROM
Device"
option for the data transfer direction.
Operation via FieldCare: using the HistoROM CONTROL parameter, select the
"HistoROM
Device" option for the data transfer direction (menu path: OPERATING
MENU
OPERATION).
Using the DOWNLOAD SELECT. parameter (OPERATION menu), select which
parameters should be overwritten.
The following parameters are overwritten depending on the option selected:
–
Configuration copy (factory setting)
All the parameters apart from the DEVICE SERIAL No., DEVICE DESIGN, PD-TAG,
DESCRIPTOR, DESCRIPTION, DEVICE ID, DEVICE ADDRESS and the parameters in the
POSITION ADJUSTMENT, PROCESS CONNECTION, SENSOR TRIM and SENSOR DATA
group.
–
Device replacement
All the parameters apart from the DEVICE SERIAL No., DEVICE ID, DEVICE DESIGN
and the parameters in the POSITION ADJUSTMENT, PROCESS CONNECTION,
SENSOR TRIM and SENSOR DATA group.
–
Electronics replace
All parameters apart from the parameters of the SENSOR DATA group.
Factory setting: Configuration copy