Deltabar S FMD77, FMD78, PMD75 with FOUNDATION Fieldbus
Commissioning
Hauser
81
Event severity
Default event category
Event source
Bit
Events in this group
High severity
Check (C)
Sensor
27
• 101: C>Sensor electronic EEPROM error
• 725: C>Sensor connection error, cycle disturbance
• 747: C>Sensor software not compatible to electronics
Electronics
26
• 703: C>Measurement error
• 704: C>Measurement error
• 705: C>Measurement error
• 737: C>Measurement error
• 738: C>Measurement error
• 739: C>Measurement error
• 742: C>Sensor connection error (upload)
• 743: C>Electronic PCB error during initialization
• 744: C>Main electronic PCB error
• 746: C>Sensor connection error - initializing
• 748: C>Memory failure in signal processor
Configuration 25
• 106: C>Downloading - please wait
• 613: C>Simulation active
Process
24
• Not used
Event severity
Default event category
Event source
Bit
Events in this group
Low severity
Out of specification
(S)
Sensor
23
• 115: S>Sensor overpressure
• 120: S>Sensor low pressure
• 715: S>Sensor over temperature
• 720: S>Sensor under temperature
• 726: S>Sensor temperature error - overrange
Electronics
22
• 717: S>Transmitter over temperature
• 718: S>Transmitter under temperature
Configuration 21
• 701: S>Adjustment outside sensor nominal range
• 710: S>Set span too small. Not allowed.
• 727: S>Sensor pressure error - overrange
Process
20
• 740: S>Calculation overflow, bad configuration
Содержание Deltabar S FMD77