Deltabar S FMD77, FMD78, PMD75 with FOUNDATION Fieldbus
Commissioning
Hauser
83
Factory setting of assignment parameters
Proceed as follows to change the diagnostic behavior of an event group:
1. Open the assignment parameter to which the group is currently assigned.
2. Change the bit of the event group from
1
to
0
. If using FieldCare, this is done via the
FF912 module by disabling the appropriate check box (see the following example).
3. Open the assignment parameter to which the group should be assigned.
4. Change the bit of the event group from
0
to
1
. If using FieldCare, this is done via by
enabling the appropriate check box (see the following example).
Example
The
Highest Severity / Electronics
group contains the events
131: Checksum error in
EEPROM: min/max segment, among others.
These should no longer be categorized as
Failure (F)
, and should be categorized as
Check (C)
instead.
1. In the FieldCare navigation window navigate to
Expert
Field Device Diagnostic
Configuration
Standard area
Configurable area
Event severity
Highest severity
High severity
Low severity
Lowest severity
Event source
1)
S
E
C
P
S
E
C
P
S
E
C
P
S
E
C
P
Bit
31
30
29
28
27
26
25
24
23
22
21
20
19
18
17
16
15...1
FD_FAIL_MAP
1
1
1
1
0
0
0
0
0
0
0
0
0
0
0
0
0
FD_CHECK_MAP
0
0
0
0
1
1
1
1
0
0
0
0
0
0
0
0
0
FD_OFFSPEC_MAP
0
0
0
0
0
0
0
0
1
1
1
1
0
0
0
0
0
FD_MAINT_MAP
0
0
0
0
0
0
0
0
0
0
0
0
1
1
1
1
0
1)
S: Sensor; E: Electronics; C: Configuration; P: Process
Содержание Deltabar S FMD77