Diagnostics and troubleshooting
Deltabar S FMD77, FMD78, PMD75 with FOUNDATION Fieldbus
94
Hauser
9.3
Diagnostic event in the operating tool
If a diagnostic event is present in the operating tool, the status signal appears in the top left
status
area along with the corresponding symbol for event behavior in accordance with NAMUR NE
107:
• Failure (F)
• Check (C)
• Maintenance required (M)
• Out of specification (S)
P01-xMx7xxxx-05-xx-xx-en-008
Calling up remedy information
1. Navigate to the "Diagnostics" menu. The diagnostic event along with text about the
event is displayed in the "Actual diagnostics" parameter.
2. On the right in the display area, hover the cursor over the "Actual diagnostics"
parameter. A tool tip with remedy information for the diagnostic event appears.
Message
Measure
Initialization, VU Electr. Defect A110
Initialization, VU Electr. Defect A114
Initialization, VU Electr. Defect A281
Initialization, VU Checksum Err. A110
Initialization, VU Checksum Err. A112
Initialization, VU Checksum Err. A171
Exchange onsite display.