Maintenance required (M)
Diagnostic
Code
Error
response
Message/description XD_
ERROR
Value Bit
BLOCK_
ERROR
Value Bit
Cause
Measure
Prio-
rity
102
Warning
M>Checksum error in
EEPROM: peakhold
segment
23
0
– Main electronics defect.
Correct measurement can
continue as long as you do not
need the peak hold indicator
function.
– Replace main electronics.
62
116 Warning
M>Download
error,
repeat download
17
0
– The file is defect.
– Use another file.
38
– During the download, the
data are not correctly
transmitted to the processor,
e.g. because of open cable
connections, spikes (ripple)
on the supply voltage or
electromagnetic effects.
– Check cable connection PC –
transmitter.
– Block off electromagnetic
effects or eliminate sources of
disturbance.
– Perform reset (code 7864)
and recalibrate the device.
– Repeat download.
134 Warning
M>EEPROM
lifetime
WARNING
17
0
– Writing too often to
EEPROM.
– Reduce write accessing to
EEPROM.
65
700 Warning
M>Last
configuration
not stored
23
0
– An error occurred when
writing or reading
configuration data or the
power supply was
disconnected.
– Perform reset (code 7864)
and recalibrate the device.
63
– Main electronics defect.
– Replace main electronics.
702
Warning
M>HistoROM data not
consistent
17
0
– Data were not written
correctly to the HistoROM,
e.g. if the HistoROM was
detached during the writing
process.
– Repeat upload.
– Perform reset (code 7864)
and recalibrate the device.
64
– HistoROM does not have any
data.
– Copy suitable data to the
HistoROM. (
Section 6.5.1 "Copying
configuration data".)
706 Warning
M>Configuration
in
HistoROM and device
not identical
23
0
– Configuration (parameters) in
the HistoROM and in the
device not identical.
– Copy data from the device to
the HistoROM. (
Section 6.5.1 "Copying
configuration data".)
– Copy data from the HistoROM
configuration data".) The
message remains if the
HistoROM and the device
have different software
versions. The message
disappears if you copy the
data from the device to the
HistoROM.
– Device reset codes such as 1
or 40864 do not have any
effect on the HistoROM. That
means that if you perform a
reset, the configurations in
the HistoROM and in the
device can be different.