Diagnostics and troubleshooting
Deltabar S FMD77, FMD78, PMD75 with FOUNDATION Fieldbus
102
Hauser
710
Warning
B>Set span too small.
Not allowed.
18
0
– Values for calibration (e.g.
lower range value and upper
range value) are too close
together.
– Adjust calibration to suit
sensor (
see Operating
Instructions BA00303P,
parameter description),
CAL_MIN_SPAN/MINIMUM
SPAN parameter).
60
– The sensor was replaced and
the customer-specific
configuration does not suit
the sensor.
– Adjust calibration to suit
sensor.
– Replace sensor with a suitable
sensor.
– Unsuitable download carried
out.
– Check configuration and
perform download again.
711
Alarm
C>LRV or URV out of
edit limits
18
0
– Lower range value and/or
upper range value undershoot
or overshoot the sensor range
limits.
– Reset the LRV and/or URV to
suit the sensor. Pay attention
to position adjustment.
37
– The sensor was replaced and
the customer-specific
configuration does not suit
the sensor.
– Reset the LRV and/or URV to
suit the sensor. Pay attention
to position adjustment.
– Replace sensor with a suitable
sensor.
– Unsuitable download carried
out.
– Check configuration and
perform download again.
713 Alarm
C>100%
POINT
(LEVEL_100_PERCEN
T_VALUE) level out of
edit limits
18
0
– The sensor was replaced.
– Carry out calibration again.
46
719 Alarm
C>Y-VAL
(TAB_XY_VALUE) of
lin. table out of edit
limits
19
0
– At least one Y-VALUE
(TAB_XY_VALUE) in the
linearization table is below
the SCALE_OUT, EU_0/
TANK CONTENT MIN. or
above the SCALE_OUT,
EU_100/TANK CONTENT
MAX.
– Carry out calibration again.
(
See Operating Instructions
BA00303P, parameter
description), Section 5).
47
721 Alarm
C>ZERO
POSITION
(LEVEL OFFSET) level
out of edit limits
18
0
– LEVEL MIN
(LINEAR_LEVEL_ MIN) or
LEVEL MAX
(LINEAR_LEVEL_MAX) has
been changed.
– Perform reset (code 2710)
and recalibrate the device.
48
722 Alarm
C>EMPTY
CALIB.
(SCALE_OUT, EU_0)
or FULL CALIB.
(SCALE_OUT,
EU_100) out of edit
limits
18
0
– LINEAR_ LEVEL_MIN/LEVEL
MIN or
LINEAR_LEVEL_MAX/
LEVEL MAX has been
changed.
– Perform reset (code 2710)
and recalibrate the device.
49/50
723 Alarm
C>MAX.
FLOW
(SCALE_OUT,
EU_100) out of edit
limits
18
0
– FLOW_TYPE/FLOW-MEAS.
TYPE has been changed.
– Carry out calibration again.
51
741 Alarm
C>TANK
HEIGHT
(LEVEL_TANK_HEIGH
T) out of edit limits
18
0
– LINEAR_LEVEL_MIN/LEVEL
MIN or
LINEAR_LEVEL_MAX/
LEVEL MAX has been
changed.
– Perform reset (code 2710)
and recalibrate the device.
52
Diagnostic
Code
Error
response
Message/description XD_
ERROR
Value Bit
BLOCK_
ERROR
Value Bit
Cause
Measure
Prio-
rity