Deltabar S FMD77, FMD78, PMD75 with FOUNDATION Fieldbus
Installation
Hauser
13
4
Installation
NOTICE
Incorrect handling!
Damage of the device!
‣
Disassembly of the screws with item number (1) is not permissible under any
circumstances and will result in loss of warranty.
A0025336
4.1
Incoming acceptance and storage
4.1.1
Incoming acceptance
• Check the packaging and the contents for damage.
• Check the shipment, make sure nothing is missing and that the scope of supply matches
your order.
4.1.2
Transport
WARNING
!
Incorrect transport
Housing and diaphragm may be damaged and and there is a risk of injury!
‣
Transport the measuring device to the measuring point in its original packaging or by the
process connection (with secure transport protection for the diaphragm).
‣
Follow the safety instructions and transport conditions for devices of more than 18 kg
(39.69 lbs).
‣
Do not use capillaries as a carrying aid for the diaphragm seals.
4.1.3
Storage
The device must be stored in a dry, clean area and protected against impact (EN 837-2).
Storage temperature range:
• –40 to +90°C (–40 to +194 °F)
• Onsite display: –40 to +85°C (–40 to +185°F)
• Separate housing: –40 to +60°C (–40 to +140°F)
4.2
Installation conditions
4.2.1
Dimensions
For dimensions, please refer to the Technical Information for Deltabar S TI00382P,
"Mechanical construction" section.