Diagnostics and troubleshooting
Hauser
103
746 Warning
C>Sensor
connection
error - initializing
20
0
– Electromagnetic effects are
greater than specifications in
the technical data.
(
See Section 10.) This
message normally only
appears briefly.
– Wait a few minutes.
– Restart the device. Perform
reset (Code 7864).
– Block off electromagnetic
effects or eliminate source of
disturbance.
28
– Overpressure or low pressure
present.
– Reduce or increase pressure.
750 Warning
C>Configuration
not
permitted
18
0
– By means of the operation
profile, options were selected
for the configuration of the
device but the options do not
suit one another. For
example, if the option "1"
(linearization table) was
selected for LIN_TYPE and
the unit "1347 (m
3
/s)" was
selected for
PRIMARY_VALUE_UNIT.
– Check configuration.
– Perform reset (code 7864)
and recalibrate the device.
53
Diagnostic
Code
Error
response
Message/description XD_
ERROR
Value Bit
BLOCK_
ERROR
Value Bit
Cause
Measure
Prio-
rity