Pressure measurement in gases
1
2
A0028473
1
Cerabar M
2
Shutoff device
Mount the Cerabar M with the shutoff device above the tapping point so that any conden-
sate can flow into the process.
Pressure measurement in vapors
1
1
2
2
3
4
A0028474
1
Cerabar M
2
Shutoff device
3
U-shaped siphon
4
Circular siphon
• Mount the Cerabar M with the siphon below the tapping point.
• Fill the siphon with liquid before commissioning. The siphon reduces the temperature to
almost ambient temperature.