Cerabar M, Deltapilot M
Description of device parameters
Hauser
77
Write permission
No write permissions. Parameter is read only.
Description
Displays the measured pressure after sensor recalibration, position adjustment and damp-
ing.
Sensor pressure (109)
Write permission
No write permissions. Parameter is read only.
Description
Displays the measured pressure before the sensor trim and position adjustment.
Corrected press. (172)
Navigation
Setup → Corrected press.
Write permission
Operators/Service engineers/Expert
Description
Displays the measured pressure after sensor trim and position adjustment.
Note
If this value is not equal to "0", it can be corrected to "0" by the position adjustment.
Pressure af. damp (111)
Write permission
No write permissions. Parameter is read only.
Description
Displays the measured pressure after sensor trim, position adjustment and damping.