Cerabar M, Deltapilot M
Operation
Hauser
39
The following table illustrates the symbols that can appear on the local display. Four sym-
bols may appear at the same time.
Symbol Meaning
A0018154
Lock symbol
The operation of the device is locked. Unlock the device, → 41.
A0018155
Communication symbol
Data transfer via communication
A0013958
Error message "Out of specification"
The device is being operated outside its technical specifications (e.g. during startup or cleaning).
A0013959
Error message "Service mode"
The device is in the Service mode (e.g. during a simulation).
A0013957
Error message "Maintenance required"
Maintenance is required. The measured value remains valid.
A0013956
Error message "Failure detected"
An operating error has occurred. The measured value is no longer valid.
6.4.1
Operating keys on the display and operating module
Operating key(s)
Meaning
A0017879
• Navigate downwards in the picklist
• Edit the numerical values and characters within a function
A0017880
• Navigate upwards in the picklist
• Edit the numerical values and characters within a function
A0017881
• Confirm entry
• Jump to the next item
• Select a menu item and activate the edit mode
A0017879
and
A0017881
Contrast setting of local display: darker
A0017880
and
A0017881
Contrast setting of local display: brighter
A0017879
and
A0017880
ESC functions:
• Exit edit mode for a parameter without saving the changed value
• You are in a menu at a selection level. Each time you press the keys simultaneously, you
go up a level in the menu.
6.4.2
Operating example: Parameters with a picklist
Example: selecting "Deutsch" as the language of the menu.
Language
000 Operation
1 English
"English" is set as the menu language (default value).
A in front of the menu text indicates the option that is currently active.
Deutsch
2
Deutsch
Select "Deutsch" with or .
English