Installation
Cerabar M, Deltapilot M
18
Hauser
Pressure measurement in liquids
1
2
3
A0028491
1
Cerabar M
2
Shutoff device
Mount the Cerabar M with the shutoff device below or at the same level as the tapping
point.
Level measurement
A0028492
• Always install the device below the lowest measuring point.
• Do not install the device at the following positions:
• in the filling curtain
• in the tank outlet
• in the suction area of a pump
• at a point in the tank which could be affected by pressure pulses from the agitator.
• The calibration and functional test can be carried out more easily if you mount the device
downstream from a shutoff device.