Description of device parameters
Cerabar M, Deltapilot M
76
Hauser
Write permission
Maintenance/Expert
Description
Select the unit for the temperature measured values.
Selection
• °C
• °F
• K
Note
The setting affects the unit for the "Sensor temp." parameter.
Factory setting
°C
Sensor temp. (110)
Write permission
No write permissions. Parameter is read only.
Description
Displays the temperature currently measured in the sensor. This can deviate from the
process temperature.
13.7 Expert → Measurement → Pressure
Set LRV (013)
Write permission
Operator/Maintenance/Expert
Description
Set the lower-range value – without reference pressure.
Enter the pressure value for the lower current value (4 mA).
Factory setting
0.0 or as per order specifications
Set URV (014)
Write permission
Operator/Maintenance/Expert
Description
Set the upper-range value – without reference pressure.
Enter the pressure value for the upper current value (20 mA).
Factory setting
Upper range limit or as per order specifications
Meas. pressure (020)