Description of device parameters
Cerabar M, Deltapilot M
80
Hauser
Prerequisite
• "Level selection" = In pressure
• "Calibration mode" = Dry -> entry
• "Calibration mode" = Wet -> display
Factory setting
URL of the sensor module
Level before lin. (019)
Write permission
No write permissions. Parameter is read only.
Description
Displays the level value prior to linearization.
13.9 Expert → Measurement → Sensor limits
LRL sensor (101)
Write permission
No write permissions. Parameter is read only.
Description
Displays the lower-range limit of the sensor
URL sensor (102)
Write permission
No write permissions. Parameter is read only.
Description
Displays the upper-range limit of the sensor.
13.10 Expert → Measurement → Sensor trim
Lo trim measured (129)
Write permission
Parameter is read only.
Only Hauser Service has write permission.
Description
Displays the reference pressure present to be accepted for the lower calibration point.
Hi trim measured (130)