Cerabar M, Deltapilot M
Description of device parameters
Hauser
81
Write permission
Parameter is read only.
Only Hauser Service has write permission.
Description
Displays the reference pressure present to be accepted for the upper calibration point.
Lo trim sensor (131)
Description
Sensor recalibration by entering a target pressure while simultaneously and automatically
accepting a reference pressure present for the lower calibration point.
Hi trim sensor (132)
Description
Sensor recalibration by entering a target pressure while simultaneously and automatically
accepting a reference pressure present for the upper calibration point.
13.11 Expert → Output → Current output
Output current (054)
Write permission
Operators/Service engineers/Expert
Description
Displays the current current value.
Output fail mode (190)
Write permission
Operators/Service engineers/Expert
Description
Select Output fail mode.
In the event of an alarm, the current assumes the current value specified with this parame-
ter.
Options
• Max: can be set from 21 to 23 mA, see also "High alarm curr."
• Hold: last measured value is held.
• Min: 3.6 mA
Factory setting
Max (22 mA)
Get URV (pressure measuring mode) (015)