Diagnostics and troubleshooting
Cerabar M, Deltapilot M
62
Hauser
10.1.2 List of diagnostic events
"C" messages
Diagnostic event
Cause
Remedial measure
Code
Description
C482 Simul. output
Simulation of the current output is switched on, i.e. the device is not measuring at present.
End the simulation
C484 Error simul.
Fault state simulation is switched on, i.e. the device is not measuring at present.
End the simulation
C485 Measure simul.
Simulation is switched on, i.e. the device is not measuring at present.
End the simulation
C824 Process pressure • Overpressure or low pressure present.
• Electromagnetic effects are greater than specifications in the technical data. This message
normally only appears briefly.
• Check the pressure value
• Restart the device
• Perform a reset
"F" messages
Diagnostic event
Cause
Remedial measure
Code
Description
F002 Sens. unknown
Sensor does not suit the device (electronic sensor nameplate).
Contact Hauser Service
F062 Sensor conn.
• Cable connection between sensor and main electronics disconnected.
• Sensor defective.
• Electromagnetic effects are greater than specifications in the technical data.
This message normally only appears briefly.
• Check the sensor cable
• Replace the electronics module
• Contact Hauser Service
• Replace the sensor (snap-on version)
F081 Initialization
• Cable connection between sensor and main electronics disconnected.
• Sensor defective.
• Electromagnetic effects are greater than specifications in the technical data.
This message normally only appears briefly.
• Perform a reset
• Check the sensor cable
• Contact Hauser Service
F083 Memory content
• Sensor defective.
• Electromagnetic effects are greater than specifications in the technical data.
This message normally only appears briefly.
• Restart the device
• Contact Hauser Service
F140 Working range P
• Overpressure or low pressure present.
• Electromagnetic effects are greater than specifications in the technical data.
This message normally only appears briefly.
• Sensor defective.
• Check the process pressure
• Check the sensor range
F261 Electronics module • Main electronics defective.
• Fault in the main electronics.
• Restart the device
• Replace the electronics module
F282 Memory
• Fault in the main electronics.
• Main electronics defective.
• Restart the device
• Replace the electronics module
F283 Memory content
• Main electronics defective.
• Electromagnetic effects are greater than the specifications in the technical
data.
• The supply voltage is disconnected when writing.
• An error occurred when writing.
• Perform a reset
• Replace the electronics module
F419 Current cycle
The Back-To-Box command is executed.
Restart the device
F841 Sensor range
• Overpressure or low pressure present.
• Sensor defective.
• Check the pressure value
• Contact Hauser Service
"M" messages
Diagnostic event
Cause
Remedial measure
Code
Description
M002 Sens. unknown
Sensor does not suit the device (electronic sensor nameplate). Device
continues measuring.
Contact Hauser Service
M283 Memory content • Cause as indicated for F283.
• Correct measurement can continue as long as you do not need the
peakhold indicator function.
• Perform a reset
• Replace the electronics module