STP-iXA2205 Series Turbomolecular Pump
Nov. 10
116
Issue 1-h
5
SERIAL
COMMUNICATION
PROTOCOL
5.4.8 ReadSetPoint
Function: Reads the setting value of the "Speed Set Point" and TMS temperature. The "Speed Set
Point" data is the same data as that of "ReadSpeedSetPoint".
Transmission frame:
PC->SIM Stx 0
0
1
?
d Etx LRC
SIM->PC
Ack
PC->SIM
Ack
SIM->PC Stx 0
0
1 Sp d
Parameter 1 to 2
Etx LRC
Parameter
Item
Data format
Remarks
1
Speed Set Point
(Unit: Hz)
16-bits hexadecimal
coded ASCII
2
TMS temperature setting
(Unit: ºC)
16-bits hexadecimal
coded ASCII
Example:
Speed Set Point
: 01F4
hex
= 500 Hz = 30,000 rpm
TMS temperature setting
: 0046
hex
= 70 ºC
Parameter
1
2
ASCII
0 1 F
4 0 0 4
6
HEX
30 31 46 34 30 30 34 36
5.4.9 ReadMotorTemp
Function: Reads the measured motor temperature.
Transmission frame:
PC->SIM Stx 0
0
1
?
e Etx LRC
SIM->PC
Ack
PC->SIM
Ack
SIM->PC Stx 0
0
1 Sp e
Parameter 1
Etx LRC
Содержание STP Series
Страница 22: ...STP iXA2205 Series Turbomolecular Pump Nov 10 12 Issue 1 h INTRODUCTION 1 This page intentionally blank ...
Страница 74: ...STP iXA2205 Series Turbomolecular Pump Nov 10 64 Issue 1 h INSTALLATION 3 This page intentionally blank ...
Страница 95: ...STP iXA2205 Series Turbomolecular Pump Nov 10 85 Issue 1 h 4 OPERATION Figure 55 X2 REMOTE output signal pins ...
Страница 98: ...STP iXA2205 Series Turbomolecular Pump Nov 10 88 Issue 1 h 4 OPERATION This page intentionally blank ...
Страница 172: ...STP iXA2205 Series Turbomolecular Pump Nov 10 162 Issue 1 h MAINTENANCE AND INSPECTION 8 This page intentionally blank ...