STP-iXA2205 Series Turbomolecular Pump
Nov. 10
114
Issue 1-h
5
SERIAL
COMMUNICATION
PROTOCOL
5.4.6 ReadVersion
Function: Read the software version.
Transmission frame:
PC->SIM Stx 0
0
1
?
V
Etx LRC
SIM->PC
Ack
PC->SIM
Ack
SIM->PC Stx 0
0
1 Sp V
Parameter 1 to 24 Etx LRC
Parameter
Item
Data format
Remarks
1 to 16
Control unit software version
8-bits hexadecimal
coded ASCII
17 to 20
Motor driver software version
4-bits ASCII character
Ver.1.2 = 0120
21 to 24
AMB software version
4-bits ASCII character
Ver.3.4 = 0340
Example:
Control unit software version
: 36335F4120312E302020202020202020
hex
= 63_A 1.0
Motor driver software version
: 0120
hex
= 1.2
DSP software version
: 0340
hex
= 3.4
Parameter
1
2
3
4
5
6
7
8
9
10
"6"
"3"
"_"
"A"
" "
"1"
"."
"0"
" "
" "
ASCII
3 6 3
3 5 F 4
1
2
0
3
1
2
E
3
0
2 0 2 0
HEX
33 36 33 33 35 46 34 31 32 30 33 31 32 45 33 30 32 30 32 30
Parameter
11
12
13
14
15
16
17 18 19 20 21 22 23 24
" "
" "
" "
" "
" "
" "
ASCII
2 0 2
0 2 0 2
0
2
0
2
0
0
1
2
0
0 3 4 0
HEX
32 30 32 30 32 30 32 30 32 30 32 30 30 31 32 30 30 33 34 30
Содержание STP Series
Страница 22: ...STP iXA2205 Series Turbomolecular Pump Nov 10 12 Issue 1 h INTRODUCTION 1 This page intentionally blank ...
Страница 74: ...STP iXA2205 Series Turbomolecular Pump Nov 10 64 Issue 1 h INSTALLATION 3 This page intentionally blank ...
Страница 95: ...STP iXA2205 Series Turbomolecular Pump Nov 10 85 Issue 1 h 4 OPERATION Figure 55 X2 REMOTE output signal pins ...
Страница 98: ...STP iXA2205 Series Turbomolecular Pump Nov 10 88 Issue 1 h 4 OPERATION This page intentionally blank ...
Страница 172: ...STP iXA2205 Series Turbomolecular Pump Nov 10 162 Issue 1 h MAINTENANCE AND INSPECTION 8 This page intentionally blank ...