STP-iXA2205 Series Turbomolecular Pump
Nov. 10
117
Issue 1-h
5
SERIAL
COMMUNICATION
PROTOCOL
Parameter
Item
Data format
Remark
1
Motor temperature
(Unit: ºC)
16-bits hexadecimal
coded ASCII
Example:
Motor temperature
: 0014
hex
= 20 ºC (68 ºF)
Parameter
1
ASCII
0 0 1
4
HEX
30 30 31 34
5.4.10 ReadStatus
Function: Reads various settings (Remote mode, TMS function, Emergency vent valve).
Transmission frame:
PC->SIM Stx 0
0
1
?
f Etx LRC
SIM->PC
Ack
PC->SIM
Ack
SIM->PC Stx 0
0
1 Sp
f
Parameter 1 to 4
Etx LRC
Parameter
Item
Data format
Remarks
1
Remote mode setting
8-bits hexadecimal
coded ASCII
Refer to Table 37
2
TMS function setting
8-bits hexadecimal
coded ASCII
00
hex
: ENABLE
Excluding 00
hex
: DISABLE
3
[System reservation]
8-bits hexadecimal
coded ASCII
4
Emergency vent valve setting
8-bits hexadecimal
coded ASCII
00
hex
: ENABLE
Excluding 00
hex
: DISABLE
Содержание STP Series
Страница 22: ...STP iXA2205 Series Turbomolecular Pump Nov 10 12 Issue 1 h INTRODUCTION 1 This page intentionally blank ...
Страница 74: ...STP iXA2205 Series Turbomolecular Pump Nov 10 64 Issue 1 h INSTALLATION 3 This page intentionally blank ...
Страница 95: ...STP iXA2205 Series Turbomolecular Pump Nov 10 85 Issue 1 h 4 OPERATION Figure 55 X2 REMOTE output signal pins ...
Страница 98: ...STP iXA2205 Series Turbomolecular Pump Nov 10 88 Issue 1 h 4 OPERATION This page intentionally blank ...
Страница 172: ...STP iXA2205 Series Turbomolecular Pump Nov 10 162 Issue 1 h MAINTENANCE AND INSPECTION 8 This page intentionally blank ...