STP-iXA2205 Series Turbomolecular Pump
Nov. 10
82
Issue 1-h
4
OPERATION
4.9.3 Rotation INHIBIT signal
When using rotation INHIBIT signal, short-circuit the "INHIBIT ENABLE IN" signal. Relations between
rotation INHIBIT signal input and pump operation state are shown in Table 23.
Signal input
Pump operation
After short-circuit of rotation INHIBIT input
signal (A), the START operation is
performed (B).
INHIBIT IN
Open
㩷
Close
START operation
ŏ
The pump accelerates when the START
operation is performed (B).
After the START operation (A), the rotation
INHIBIT input signal is short-circuited (B).
INHIBIT IN
Open
㩷
Close
START operation
ŏ
The pump does not accelerate when the
START operation is performed (A).
ŏ
The pump accelerates when the rotation
INHIBIT input signal is short-circuited (B).
The rotation INHIBIT input signal is opened
(A) during acceleration or normal
operation.
INHIBIT IN
Open
㩷
Close
START operation
< MANUAL/REMOTE operation mode >
ŏ
The pump decelerates and stops when the
rotation INHIBIT input signal is opened (A).
ŏ
After the operation (A), the pump continues
deceleration and stops even by
short-circuiting the rotation INHIBIT input
signal (B). However, when the START signal
is input, the pump will accelerate.
Table 23 - Rotation INHIBIT signal input
A B
A B
A B
Содержание STP Series
Страница 22: ...STP iXA2205 Series Turbomolecular Pump Nov 10 12 Issue 1 h INTRODUCTION 1 This page intentionally blank ...
Страница 74: ...STP iXA2205 Series Turbomolecular Pump Nov 10 64 Issue 1 h INSTALLATION 3 This page intentionally blank ...
Страница 95: ...STP iXA2205 Series Turbomolecular Pump Nov 10 85 Issue 1 h 4 OPERATION Figure 55 X2 REMOTE output signal pins ...
Страница 98: ...STP iXA2205 Series Turbomolecular Pump Nov 10 88 Issue 1 h 4 OPERATION This page intentionally blank ...
Страница 172: ...STP iXA2205 Series Turbomolecular Pump Nov 10 162 Issue 1 h MAINTENANCE AND INSPECTION 8 This page intentionally blank ...