STP-iXA2205 Series Turbomolecular Pump
Nov. 10
87
Issue 1-h
4
OPERATION
4.10
Operating the TMS unit (for use with the TMS unit)
4.10.1 Before starting
Check the following items before starting:
1. Check that the TMS heater cable is securely connected to the power unit.
2. Check that the TMS valve, the cooling water pipe, and the TMS connection cable are securely
connected.
4.10.2 Starting/Stopping
Power on the STP pump to automatically start the TMS unit.
Power off the STP pump to automatically stop the TMS unit.
When STP pump is in the power ON state, the TMS unit operates regardless of the status of the STP
pump. When the STP pump is the cause of the malfunction, the TMS unit is stopped.
4.10.3 Setting the TMS unit function
When the temperature of the STP pump base is lower than the setting value, the TMS heater is turned
"ON", and the TMS valve is turned "OFF" to heat the STP pump. When the temperature of the STP
pump base is higher than the setting value, the TMS heater is turned "OFF", and the TMS valve is
turned "ON" to cool the STP pump.
Содержание STP Series
Страница 22: ...STP iXA2205 Series Turbomolecular Pump Nov 10 12 Issue 1 h INTRODUCTION 1 This page intentionally blank ...
Страница 74: ...STP iXA2205 Series Turbomolecular Pump Nov 10 64 Issue 1 h INSTALLATION 3 This page intentionally blank ...
Страница 95: ...STP iXA2205 Series Turbomolecular Pump Nov 10 85 Issue 1 h 4 OPERATION Figure 55 X2 REMOTE output signal pins ...
Страница 98: ...STP iXA2205 Series Turbomolecular Pump Nov 10 88 Issue 1 h 4 OPERATION This page intentionally blank ...
Страница 172: ...STP iXA2205 Series Turbomolecular Pump Nov 10 162 Issue 1 h MAINTENANCE AND INSPECTION 8 This page intentionally blank ...