![Edwards STP Series Скачать руководство пользователя страница 41](http://html.mh-extra.com/html/edwards/stp-series/stp-series_instruction-manual_2366164041.webp)
STP-iXA2205 Series Turbomolecular Pump
Nov. 10
31
Issue 1-h
INSTALLATION
3
When installing the STP pump to the vacuum equipment in upside-down position, the STP-
iXA2205CP(V)/CIP can be loft with the eyebolts attached to the dedicated legs.
㸦
see Figure 21
㸧
A crane and eyebolts due to lifting operations should withstand the load of five times or more the
weight of the STP pump, and rope should be seven time or more. Use an eyebolt which confirms to
ISO03266. When lifting the pump, use 2 or more ropes.
Figure 21 - STP pump with eyebolts
Eyebolt