STP-iXA2205 Series Turbomolecular Pump
Nov. 10
108
Issue 1-h
5
SERIAL
COMMUNICATION
PROTOCOL
Parameter
Item
Data format
Remark
1
Pump operation command 8-bits hexadecimal coded
ASCII
Refer to Table 33
Pump operation command
Value
START
1
STOP
2
RESET
4
Table 33 - Pump operation commands
Example:
Pump operation command : RESET operation = 4 = 04
hex
Parameter
1
ASCII
0 4
HEX
30 34
5.4.4 ReadFailMess
Function: Reads the errors being detected. This data is the same data as that of "ReadModFonct"
parameter 2 to 81 or "ReadModFonctWithWarning" parameter 3 to 82.
Transmission frame:
PC->SIM Stx 0
0
1
?
F Etx LRC
SIM->PC
Ack
PC->SIM
Ack
SIM->PC Stx 0
0
1 Sp F
Parameter 1 to 81
*1
Etx LRC
Содержание STP Series
Страница 22: ...STP iXA2205 Series Turbomolecular Pump Nov 10 12 Issue 1 h INTRODUCTION 1 This page intentionally blank ...
Страница 74: ...STP iXA2205 Series Turbomolecular Pump Nov 10 64 Issue 1 h INSTALLATION 3 This page intentionally blank ...
Страница 95: ...STP iXA2205 Series Turbomolecular Pump Nov 10 85 Issue 1 h 4 OPERATION Figure 55 X2 REMOTE output signal pins ...
Страница 98: ...STP iXA2205 Series Turbomolecular Pump Nov 10 88 Issue 1 h 4 OPERATION This page intentionally blank ...
Страница 172: ...STP iXA2205 Series Turbomolecular Pump Nov 10 162 Issue 1 h MAINTENANCE AND INSPECTION 8 This page intentionally blank ...