Compression Test Setup
Set the test mode to compression (
Settings
>
Test Mode
>
Compression
). This setting is specific to
the test template being used and can be saved to a template.
Vertical Configuration
1. Install the desired load cell (see
Appendix B
) to the Motor Bracket B of the UniVert
2. Attach the compression platen directly to the measuring end of the load cell. Choose an
M5 or M3 platen depending on the load cell.
Finger tighten all components. Beware of exerting excessive off-axis and torquing forces
to avoid damage to the load cell.