38
Figure 5: Tracking to 33% strain with overlaid force graph.
Exporting Tracked Data
Tracked grid point coordinates and calculated grid
strains can be exported to a .csv file. From the Export
menu, select
Data Points
to display the dialogue shown
on the right. The image list is automatically set to the
currently selected images in the test image list tree.
This image list can be changed by selecting new
images in the test image list tree and pressing the set
button. When exporting strains data from
Selected
Tracking Points
, all four points of a grid cell need to be
selected for data from that cell to be exported.
Exported data is labeled with associated ID numbers
(grid point ID numbers for exported coordinates and
grid cell ID numbers for exported strain values). To
display ID numbers on tracking grids, see Image
Tracking: The
Points Display Option
. In the following
example (Figure 6), grid point ID numbers are shown in
yellow and grid cell ID numbers are shown in white.
The exported .csv file is written to the test directory.
.
Содержание UniVert
Страница 1: ...UniVert Mechanical Test System User Manual version 3 0 ...
Страница 36: ...UniVert User Manual 31 ...
Страница 44: ...UniVert User Manual 39 Figure 6 Tracking Grid showing ID Numbers Figure 7 Sample Exported Data ...
Страница 78: ...UniVert User Manual 73 ...
Страница 79: ...UniVert User Manual 74 ...
Страница 80: ...UniVert User Manual 75 ...
Страница 81: ...UniVert User Manual 76 ...
Страница 82: ...UniVert User Manual 77 ...