UniVert User Manual
46
7. Gently place a test weight onto the platen. This weight should be between
25%-90%
of
the load cell full scale value to ensure accurate calibration. Centre the weight on the load-
bearing surface to ensure in-line loads and use weights corresponding to the load cell
capacity range.
Do not overload the load cell.
8. Take a measurement of the calibration test weight.
c. Enter the weight in grams in the
Enter Weight
dialogue
d. Click
Read
9. If desired, repeat steps 7 and 8 for multiple weights covering most of the load cell range.
However, at least one weight of at least 50% of the load cell range is recommended.
10.
Click “Run”.
11. Check the
Relative Calibration Factor
and
Absolute Calibration Factor
for reasonable
values. The Relative Calibration Factor is a ratio of the new calibration factor to the
previous calibration factor. Ideally, this value should be between 0.99-1.01. The
Absolute Calibration Factor is the actual calibration factor in mV/V. This value should be
between 18-45.
12.
Click “Apply”
13.
Click “Done”
Note: Only take one reading per calibration weight used. Multiple readings of a single load
value can cause an error in the calibration algorithm
Be sure to zero
the load cell again once the UniVert is in the desired testing
configuration
Содержание UniVert
Страница 1: ...UniVert Mechanical Test System User Manual version 3 0 ...
Страница 36: ...UniVert User Manual 31 ...
Страница 44: ...UniVert User Manual 39 Figure 6 Tracking Grid showing ID Numbers Figure 7 Sample Exported Data ...
Страница 78: ...UniVert User Manual 73 ...
Страница 79: ...UniVert User Manual 74 ...
Страница 80: ...UniVert User Manual 75 ...
Страница 81: ...UniVert User Manual 76 ...
Страница 82: ...UniVert User Manual 77 ...