UniVert User Manual
25
Range Limits
Range limits can be accessed by selecting
Range Limits
from the Settings menu.
Soft Range Limits
To avoid collisions or sample destruction, it is useful to assign reasonable travel limits to the
actuators. This is especially important during force control or preloading, where if specified loads
cannot be reached, the actuators will keep moving until they reach the soft limits.
•
Min and Max Displacement
, in mm, set the minimum and maximum positions to which
the grips or platens can move. These settings require adjustment especially when the
fixtures are changed.
•
Max Force
, in N, sets the maximum value of force at which a test will be automatically
stopped. The default value is set to the capacity of the load cells. Setting this value may
be useful to prevent sample destruction or overloading of the load cells.