UniVert User Manual
22
Advanced System Settings Dialogue
Advanced Settings are available by selecting
Advanced
from the
Settings
menu. All these
settings are specific to the test template selected.
Data Averaging
The load cells are sampled at 100Hz with hardware using a central moving average of 8 samples.
Further data averaging can be applied to smooth noisy data by selecting Apply Averaging to
Force Data) and selecting a further central moving average.