UniVert User Manual
53
If system stiffness compensation is enabled, a calibration must be performed to determine the
correct compensation factor. This factor depends on the load cell being used and a new
calibration must be completed each time the load cell is changed. There is a software utility to
automatically perform this calibration. Before starting the software utility, attach the compression
platens to the UniVert system. Also, place the Stiffness Compensation spacer that is provided (a
small washer) onto the exact center of the lower platen.
After placing the calibration spacer, slowly jog the platens together until there is a very small gap
between the upper platen and the spacer. If this gap is too large, the software utility will take a
very long time to run. However, do not allow the upper platen and the spacer to come in contact.
If the platens crash together the load cell will likely be damaged.
UniVert Tip: Jog Slowly to Protect Load Cells
Jog the platens together at a slow speed to prevent crashing them together and damaging the
load cell. Use lower jog speeds the closer the platens are to each other.
Содержание UniVert
Страница 1: ...UniVert Mechanical Test System User Manual version 3 0 ...
Страница 36: ...UniVert User Manual 31 ...
Страница 44: ...UniVert User Manual 39 Figure 6 Tracking Grid showing ID Numbers Figure 7 Sample Exported Data ...
Страница 78: ...UniVert User Manual 73 ...
Страница 79: ...UniVert User Manual 74 ...
Страница 80: ...UniVert User Manual 75 ...
Страница 81: ...UniVert User Manual 76 ...
Страница 82: ...UniVert User Manual 77 ...