UniVert User Manual
12
Step 3: Specify Test Type
Access
Test Mode
in the
Settings
menu to select the appropriate test mode and install the
platens, grips, or 3-point bend attachments (see
Appendix D
for this procedure).
Step 4: Zero the Load Cell (occasionally)
While it is not necessary to zero the load cell with every test sequence, we suggest zeroing the
load cell at the start of a new test session. With repeated use, the zero point of the load cell can
drift. By zeroing the load cell, you are ensuring that the force measurements are accurate. The
force offset can also be affected by changing the fixtures.
To zero the load cell, select
Zero Load Cell
from the Tools menu or press
on the toolbar.
System Alert
You should zero the load cell after attaching a platen or grip but before mounting a sample. If
a specimen is already loaded when the load cell is zeroed, the force reading will include an
offset load.