UniVert User Manual
44
System Calibration and Advanced Tools
Load Cell Calibration
The load cells are preset with a calibration factor during manufacturing and quality check.
However, calibration in the field is recommended. When the load cells are set up for the first time,
changed, or moved it is advisable to perform a load cell calibration. This process will require a set
of calibration weights (not included with system). Load cells can be calibrated attached to the
UniVert system or with the Calibration Plates.
Attached to System
1. Attach the load cell to the fixed end of the UniVert (i.e. on the baseplate) with an M5x16
SHCS and 4mm hex key. Ensure the measuring end of the load cell is facing away from
the system (See
Appendix B
).
2. Affix the calibration platen to the load cell.