![CellScale UniVert Скачать руководство пользователя страница 28](http://html.mh-extra.com/html/cellscale/univert/univert_user-manual_2580705028.webp)
UniVert User Manual
23
Force Control Settings
Force Control Settings affect the rate and stability at which forces are achieved in force-controlled
procedures. These settings apply to preloads as well as Force Control Modes.
Velocity
(1-10 integer values) matches actuator jog speed. A default value of 8 is set in the
sample templates provided. Typically for stiffer or smaller specimens (i.e. requires a short
displacement to achieve desired force), use a lower velocity setting as this will allow for smoother
force controlled protocols. It may take a few trials to optimize your test protocols and settings for
your experiment.
Velocity Jog Integer
mm/s
1
0.039063
2
0.078125
3
0.15625
4
0.3125
5
0.625
6
1.25
7
2.5
8
5
9
10
10
20
UniVert Tip: Testing Force Control Settings
It is important to have appropriate Force Control Settings that match the materials you are
testing.
To safely test Force Control Settings: press
on
the tool bar, or from the Tools menu select Move
Actuators to Specified Load. The dialogue shown on
the right will appear. The UniVert will attempt to
achieve the specified load (force) but will stop when
the specified load is reached, the actuator has
displaced by the maximum specified amount, or the
timeout time has elapsed.
If the actuator oscillates around the specified loads
but never achieves them try decreasing the velocity.
If the actuator reaches the specified load too slowly,
try increasing the velocity.
Содержание UniVert
Страница 1: ...UniVert Mechanical Test System User Manual version 3 0 ...
Страница 36: ...UniVert User Manual 31 ...
Страница 44: ...UniVert User Manual 39 Figure 6 Tracking Grid showing ID Numbers Figure 7 Sample Exported Data ...
Страница 78: ...UniVert User Manual 73 ...
Страница 79: ...UniVert User Manual 74 ...
Страница 80: ...UniVert User Manual 75 ...
Страница 81: ...UniVert User Manual 76 ...
Страница 82: ...UniVert User Manual 77 ...