48
System Alert: Load Cell Handling
It is important to handle load cells carefully. Here are some things to consider while handling a
load cell:
•
The load cells are designed to handle overloads of up to 120% full scale value.
Overloading a load cell past this rating will cause permanent damage or failure.
•
Avoid applying torque and transverse forces to the load cells. Over tightening grips or
platens onto the active end of the load cell may damage it.
•
Accidentally crashing the platens or grips together will produce more force than most
load cells can handle. Be careful and reduce speed when jogging grips close
together.
Generally, avoid dropping, tapping and exerting unnecessary force to increase the life of the
load cell.