To perform system stiffness compensation calibration, follow these steps:
1. Open the System Stiffness Compensation utility by selecting Advanced
→
System
Compensation
from the Tools menu.
2. Zero the load cells.
3. Use the jog buttons to decrease the gap above the spacer. Do not contact the top of the
calibration spacer.
4. Click Run to start the calibration and set the stiffness compensation value.
5. Press OK to Reset Actuators and complete the calibration.
6. The calculated system stiffness compensation value will be displayed in the message
window.
The calibration utility will slowly drive the platens together until the full-scale load is reached on
the load cell. Resulting load displacement data is typically very linear and a best-fit slope is used
to determine the calibration factor. Compensation is calculated in compression but is assumed to
act linearly in both tension and compression.