UniVert User Manual
57
Troubleshooting
Communication Errors: Read File or Write File
When communication errors occur the live message window will display a
ReadFile
or WriteFile
failed message. Communication errors can occur if:
•
The UniVert is not powered up.
•
USB cables between the computer and UniVert are disconnected.
•
The supply power surges or is interrupted.
When a communication error occurs, close UniVert software, check connections, power cycle the
UniVert Device, and restart the UniVert software. Be sure to reset the actuators before setting up
a new test.
UniVert Tip: Preventing Power Disruptions
We recommend plugging the UniVert (and host computer) into an uninterruptable power
supply (UPS) or battery back-up.