UniVert User Manual
24
Force and Displacement Control Settings
Acceleration
(1-10 integer values). A default value of 7 is set in the sample templates provided.
Changes made to the acceleration will affect tests under both force and displacement control.
Reducing this value may help smooth force-controlled tests under low loads or displacement-
controlled tests at high speeds. Tests run with lower accelerations will have softer transitions at
direction or speed changes but may take longer than specified to achieve specified testing cycles.
Acceleration
value
m/s2
Time (s) to
achieve 20mm/s
Travel (mm) to
achieve 20mm/s
1
0.004
5.0000
50.0000
2
0.008
2.5000
25.0000
3
0.016
1.2500
12.5000
4
0.032
0.6250
6.2500
5
0.064
0.3125
3.1250
6
0.128
0.1563
1.5625
7
0.256
0.0781
0.7813
8
0.512
0.0391
0.3906
9
1.024
0.0195
0.1953
10
2.048
0.0098
0.0977
System Stiffness Compensation
System Stiffness Compensation can be used to compensate displacement readings that occur in
the UniVert system during testing typically at high loads with low displacements. Under most
loading conditions these displacements are negligible and system stiffness compensation is not
required.
If you find this is necessary for your testing, please see the
System Stiffness Compensation
section for more details and steps on how to complete the calibration procedure
.
Содержание UniVert
Страница 1: ...UniVert Mechanical Test System User Manual version 3 0 ...
Страница 36: ...UniVert User Manual 31 ...
Страница 44: ...UniVert User Manual 39 Figure 6 Tracking Grid showing ID Numbers Figure 7 Sample Exported Data ...
Страница 78: ...UniVert User Manual 73 ...
Страница 79: ...UniVert User Manual 74 ...
Страница 80: ...UniVert User Manual 75 ...
Страница 81: ...UniVert User Manual 76 ...
Страница 82: ...UniVert User Manual 77 ...