Horizontal Configuration
1. Install the load cell (see
Appendix B
) to the fixed end of the UniVert with an M5x16
SHCS using a 4mm hex key. Use the mounting hole closest to the power switch.
2. Install the grip to the Tension Load Cell Bracket
•
For an M5 threaded load cell, use the stainless steel M5 grip, M5 Tension Load
Cell bracket, M5x12 SHCS, 4mm hex key and M5 Locking Disc.
•
For an M3 threaded load cell, use the plastic M3 grip, M3 Tension Load Cell
Bracket, M3x12 SHCS, 2.5mm hex key, and M3 Locking Disc.