UniVert User Manual
119
3. Attach the M5 stand off to the baseplate of the UniVert with an M5x16 SHCS using 4mm
hex key. This may be easier to do with the UniVert in the horizontal configuration.
4. Attach the M5 locking disc and M5 grip to the M5 stand off.
5. Align the grips to be parallel by loosening the M5 screw/locking disc and rotating the load
cell assembly/grip. Only handle the load cell assembly by the coupling or the sides of the
loadcell. Twisting the grip while it is attached to the load cell can result in overloading.
6. A Zero Position Calibration is required after changing grips or platens. See the
System
Calibration and Advanced Tools
section of this manual for more details.