![CellScale UniVert Скачать руководство пользователя страница 122](http://html.mh-extra.com/html/cellscale/univert/univert_user-manual_2580705122.webp)
UniVert User Manual
117
Tension Test Setup
The low force grips are required to be used with the low force load cells. Replacement springs of
varying stiffness are provided with the low force grips. The light, medium and heavy springs will
provide
12N
,
24N
, and
60N
of gripping force, respectively.
1. Set the test mode to Tension (
Settings
>
Test Mode
>
Tension
). This setting is specific to
the test template being used and can be saved to a template.
2. Attach the M5 to M3 stand off to the non-measuring end of the low force load cell
Содержание UniVert