UniVert User Manual
50
2. Place the provided calibration spring on the bottom platen or grip. For tests in
horizontal configuration, place the UniVert in the vertical position to complete the zero
position calibration.
3. Zero the load cell using the
Zero Load Cells
button
4. Specify the spring length (mm). This figure is provided for each spring included in the
UniVert system.
5. Using the jog buttons, move the actuator such that there is a small gap between the
top of the spring and the bottom of the upper platen or grip. Do not contact the top of
the spring. Using a slow jog speed (below 5) is advisable for this reason.
6. Select
Run
and let the calibration utility proceed. The software will calibrate the zero
position automatically.
7. Press OK to Reset Actuators and complete the calibration.