8. Secure the temperature sensor to the top of the chamber using an M4x8 Flat head
fastener and the supplied D-clip
9. To activate the fluid chamber heater, the unit must be powered on and the heater cable
must be connected (the computer and software do not need to be on yet). At this point,
the heater will warm the fluid to the last used temperature set point. The temperature set
point can be changed by selecting Hardware from the Settings menu.
10. Typically, the heater will take 30 to 60 minutes to warm a full fluid chamber from room
temperature to 37°C and will take up to 90 minutes to stabilize at this temperature.