Lynx Genie 2000 Acquisition Window Adjust
Lynx DSA User's Manual - 9240227J
51
TRP Inh. Mode
Selects the TRP Inhibit Mode to
Auto
or
Manual
. In
Auto
the signal processing is
gated off for the greater of the external RESET signal OR the Internal Inhibit Time
which is shaping-dependent. In
Manual
mode, the signal processing is inhibited for
the greater of the user-selected TRP Inhibit Time OR the external RESET signal OR
the Internal Inhibit Time.
Inp. Polarity
Sets the detector signal polarity for the ENERGY input as
Positive
or
Negative
.
High Voltage Parameters
The High Voltage controls are accessible only from a PHA datasource. The
datasource type is defined during the MID Editor or MID Wizard session by selecting
MCA Number 1 for PHA or Number 2 for MCS.
The High Voltage control in the Adjust dialog provides access to a set of parameters
required to control the High Voltage Power Supply (HVPS) in the Lynx.
Figure 24: Adjust Screen’s HVPS Settings
Range
The Lynx supports six voltage ranges, + 200 volts, +1500 volts, +5000 Volts, 200
volts, -1500 volts and -5000 volts.
The 1500 and 5000 volts ranges have separate rear-panel SHV connectors for
positive and negative voltages. The 200 volts range has a single connector for both
positive and negative voltages.
Voltage Limit
The Voltage limit control establishes the HVPS’s maximum output voltage within the
selected range. The Voltage Limit setting must be established before the Voltage
control is adjusted.